MPRF Temperature Data Logger - Specifications
Operating Parameters
| Operating temperatures Radio operating temperature |
-80ºC to +400ºC (-112ºF to +752ºF) |
| Temperature accuracy | -80ºC to -40ºC (-112ºF to -40ºF): ±0.3ºC (±0.5ºF) -40ºC to +140ºC (-40ºF +284ºF): ±0.1ºC (±0.2ºF) +140ºC to +400ºC (+289ºF to +752ºF): ±0.5ºC (±0.9ºF) |
| Operating pressure | Up to 10 BAR |
Wireless
| Range | 100 feet (30 meters) Line of Sight |
| Radio Frequency | 2.4 GHz range ISM Band |
Physical
| Size | Width: 0.89 in. (22mm) X Length: 3.6 in. (91mm)** |
| Weight | 3.1 oz. (89 grams)** |
| Material | 316 stainless steel and ULTEM |
**Size and weight are shown for Tracer body only.
Monitoring Parameters
| Memory | 8,000 time/temperature observations |
| Sample rates | Operator programmable ( • 5 sec. to 10 min. interval with radio) ( • 1 sec. to 10 min. interval without radio ) |
Mechanical
| Calibration | Factory calibrated (NIST traceable) |
| Battery type | AA Lithium (field-replaceable) |
| Battery life | 6 months to 1 year (depending on usage) |
| Clock accuracy | 10 seconds per 24 hours (25ºC) |
| Certifications | • ATEX (for intrinsic safety) |