Mesa Home>DataTrace>Frequently Asked Questions

Frequently Asked Questions

General

How do I determine what firmware version is installed in my Tracers?

How can I estimate the battery life remaining in my Tracer?

Will operating a Tracer outside of its specified operating range have any negative effects on the Tracer?

Why did a Tracer I sent for repair return with a different Serial Number?

Can I use a mercury and glass thermometer to verify my Tracer?

What type of battery do I need for my DataTrace equipment?

Can I write on, or put a sticker on, my Tracer?

I just connected my MPIII PC Interface and it does not seem to work, what's wrong?

What is my Tracer's clock accuracy and how is it checked at the factory?

Is there a test that can be performed to determine if an FRB battery is being conditioned properly?

Is Mesa Laboratories ISO certified?

Is DataTrace equipment rated Intrinsically Safe and by which agencies?

Does DataTrace equipment carry a CE mark?

Temperature Tracer

Can I use boiling water (or an ice bath) to verify my Tracer?

Can the Tracer probe tip be made shorter than 0.9" (1.0" for MPIII)?

Can the MPIII Tracer be used in a microwave?

What is the difference between a thermocouple, an RTD, and a Thermistor?

Humidity Tracer

How can I clean the humidity sensor on my Humidity Tracer?

Why are the temperature values of my Humidity Tracer different from my Temperature Tracer?

Pressure Tracer

Can my Pressure Tracer work in a vacuum?

How can I clean the pressure sensor on my Pressure Tracer?

How quickly does my Pressure Tracer respond to pressure changes?

Why are the temperature values of my Pressure Tracer different from my Temperature Tracer?

General

Back To The Top

How do I determine what firmware version is installed in my Tracers?

The Tracer firmware is the internal program resident in each Tracer that tells the unit who it is and how to react. The firmware version can be determined for all Tracers with the DataTrace for Windows (DTW) software (version 3.40 or later), except those with the earliest versions of the electronics, commonly called Hybrids.

To obtain the firmware version, perform a Test Tracer on the unit. When successfully completed click the YES button when asked to view the DataTrace TRACER STATUS REPORT. The status report will appear on the screen. At the bottom of the first section in the Status Report the firmware version will appear as an alpha character.

Back To The Top

How can I estimate the battery life remaining in my Tracer?

Battery life is extremely difficult to estimate and is not a precise value. Batteries have a very wide variation in life cycles from one battery to another of the same type. In our experience battery life in general is very much affected by three application-driven situations:

1. How frequently the Tracer is used,
2. The temperature extremes and the duration of those extremes to which the Tracer is exposed, and
3. The Sample Interval defined.

In the years of experience that we have had with the Standard Temp battery (Micropack, Flatpack, and FRB Tracers), we have found that on average users receive between 9 months and 12 months of useful battery life. Obviously, this will vary, but appears to be a reasonable estimate. This corresponds reasonably well to our on-going testing in-house in our own sterilizer.

The pressure Tracer will consume, normally, twice as much power as the Temperature-only Tracer, while the Humidity Tracer is between these extremes.

The MPIII Tracer uses a considerably smaller size battery which will have a much shorter life expectation. Based on initial process testing and our normal on-going testing, we expect the battery life for MPIII Tracers to be approximately 2 to 3 months with similar caveats as above. However, the MPIII Tracers have a battery life indicator that displays an estimated remaining battery life when the Tracer is tested or Programmed.

Back To The Top

Will operating a Tracer outside of its specified operating range have any negative effects on the Tracer?

Operating a Tracer outside its specified range may affect the Tracer in several ways. It may stress the battery and/or electronic components resulting in changes in accuracy, stability, or complete failure. The impact obviously worsens the more outside the specified range to which the Tracer is exposed. While some of these changes are reversed when the Tracer is returned to the appropriate range, others may be permanent or irreparably damage the Tracer.

Back To The Top

Why did a Tracer I sent for repair return with a different Serial Number?

There are two reasons that this could happen; which reason depends on the model of Tracer being serviced.

1. MPIII Tracers will change Serial Numbers for a different reason. If the MPIII probe is damaged or needs to be replaced, the Tracer will be returned with a different Serial Number. The design of the MPIII Tracer includes the probe as part of the body assembly, unlike the earlier Tracers where they were separate items. The Serial Number is laser- etched onto the MPIII body before assembly. During service for probe damage, a new body (with a new Serial Number) must be used.

2. Older Tracers (those built before 1996) had a circuit type called a hybrid. They had serial numbers which were either MTSxxxxx or STxxxxx. While state-of-the-art at the time they were developed, they consumed sufficient power that there was a need to design a magnetic switch to prevent battery depletion while not in use. The switch was activated by the presence of a magnet in the Tracer's Storage Module.

In 1996 we introduced a new circuit design, using the latest low power microelectronic components. This newer circuit uses so little power that it is unnecessary to turn it off when not in use. The new circuit uses less power both while collecting data and while not in use. In order to distinguish which circuit design is inside a Tracer we add a leading zero after the ST prefix of the serial number (ST0xxxxx) when the newer circuitry is retrofitted. While the serial number appears in a different format, the xxxxx should be the same as the earlier Tracer serial number.

Back To The Top

Can I use a mercury and glass thermometer to verify my Tracer?

The best way to use a glass thermometer as an accurate measurement instrument is to 1) only do so in a well-stirred liquid medium, 2) only use it at its rated immersion depth, and 3) do not use it at temperatures too far from ambient.

Mercury and glass thermometers can be accurate if they are used in the situations and environments they were designed for. Many glass thermometers are rated as full immersion; this means that to achieve their rated accurately they must be fully immersed in the environment they are measuring, frequently this is air. These types of thermometers are appropriate for environmental measurements, not for use in liquid mediums.

There are glass thermometers that are calibrated and designed for measurements in liquid. However, it is even more critical that immersion depth be observed with these units. As suggested above all temperature measuring devices have a rated immersion depth, at which there will be no measurable stem effect from heat being conducted to or from the ambient environment along the stem to or from the tip. Failure to adhere to the immersion depth requirement will introduce significant inaccuracies into the measurement.

Finally, the further from ambient that the target temperature is, the more errors can creep into the measurement due to procedural or equipment related issues.

Back To The Top

What type of battery do I need for my DataTrace equipment?

The batteries used in DataTrace equipment depend on the type of Tracers being considered. Except for the non-MPIII PC Interface, which uses a standard 9 volt battery, Tracer batteries are not standard off-the-shelf designs. The battery chemistry is Lithium Thionyl Chloride (non-MPIII Tracers) or Carbon Monoflouride Lithium (MPIII Tracers) customized for use in DataTrace Tracers. These batteries are available from DataTrace along with the accessories necessary to accomplish a field battery replacement.

Back To The Top

Can I write on, or put a sticker on, my Tracer?

Some customers have unique "tags" that they wish to place on their Tracers. In some cases these reflect mapping IDs used as visual cues for validation process locations and for others they define individual pieces of capital equipment.

DataTrace recommends that no stickers be placed on Tracers for several reasons. First, it is possible that the sticker placement could conflict with Tracer communication. Second, when a Tracer is returned for service the fixtures used on the Tracer will damage or destroy the sticker during disassembly. Finally, the sticker and/or sticker adhesive could contaminate the silicone oil in the calibration bathes.

While writing on the case with an indelible ink can be helpful, it is limited. Consideration must also be given to any contamination to the process caused by the ink itself.

An alternative worth considering might be electro-chem etching. This would be safe, clean, and would add no conflict with communication or factory fixturing. This process has been used for years by DataTrace. Here is the URL to Electro-Chem Etch Metal Markings Inc., the web site of the company who manufactures our etching equipment.

http://www.ecemmi.com/

Back To The Top

I just connected my MPIII PC Interface and it does not seem to work, what's wrong?

The MPIII PC interface gets power from the serial port of the PC. However, because the optical emitters use more power for very brief periods than the PC can supply, the Interface stores energy in very large capacitors built into its circuitry. These capacitors make a total of 2 Farads, much more energy storage than conventional capacitors. Unlike batteries, these capacitors can charge and discharge over and over without wearing out, so they should never need to be replaced. But like batteries, their charge rate must be limited, and as a result it takes a little time for them to rise to operating voltage. Normally, the PC Interface is fully charged before it leaves the factory.

The Interface is being charged when it is connected to the computer and the DTW program is active. Once they are charged from zero, they do not decay much over time and usage and can normally be used right away. However, their charge could decay slightly over time due to stray current leakage. For this reason, if the interface does not work when it is first plugged into the serial port, usually waiting about 2 minutes with DTW active will bring the Interface to full power and it will start working. Even if the capacitors are fully discharged, it only takes about 15 minutes before it will work properly.

A fully charged interface can be stored for several weeks before the charge decays enough that it will not work immediately when it is connected to a PC. However, this will vary from one unit to the next.

Back To The Top

What is my Tracer's clock accuracy and how is it checked at the factory?

The Tracer does not have a clock, but uses a timing circuit, based on a highly accurate oscillator (approximately 30ppm). As with all oscillators, high temperature affects them. However, in the case of the oscillator used in your Tracer, the actual impact is not significant. For example, in the worst case where a Tracer was operating at 150°C for 24 hours, the timing would be off less than 30 seconds.

Timekeeping accuracy is evaluated inherently as part of the calibration process for each Tracer.

Initially, during the calibration procedure, the Tracer is programmed on the computer which is used to operate the calibration bath and log the reference temperatures. Following programming, the Tracer is immersed in the bath and the computer commands the bath to go to a certain temperature. When the bath software detects that the bath temperature is stable enough, measurements are taken with the reference thermometers. After the measurements are recorded, the bath is commanded to change to the next calibration temperature.

This procedure is repeated every 5°C throughout the Tracer's temperature measurement range. After the calibration data is acquired from the reference thermometers during the calibration process, the Tracer is removed from the bath, cleaned, and read.

The data that the computer acquired during calibration includes the temperature at each stable point, and the exact time these data were taken. This reference data is then compared to the readings from the Tracer, at each point in time. If the Tracer timekeeping were not accurate, it could not pass the calibration or calibration accuracy verification. Even if the Tracer calculated the proper temperature, if the point in time were off even a few seconds during this 7 to 9 hour process, the wrong temperature would be compared to the reference temperature and the Tracer would fail.

Back To The Top

Is there a test that can be performed to determine if an FRB battery is being conditioned properly?

If the FRB battery conditioning procedures are followed per the instructions, the batteries will be properly conditioned. The following procedure was suggested by our Quality Assurance Department for those wishing to monitor the battery conditioning process:

When conditioning a battery for an FRB Tracer, connect a digital voltmeter to the battery leads, and then connect the depassivation clip. Observe the voltage drop as the internal resistance of the battery's passivation layer prevents much current from flowing. Often it can start out as low as 0.5 volts when the clip is first connected. Over the next few minutes, you can watch the voltage increase as the passivation layer is eliminated and the battery's resistance decreases. For standard temp batteries (the rectangular ones), after 3-5 minutes, the voltage should climb to 2.4 volts or better. Low temp batteries (the round ones) usually take a little less time. Disconnect the depassivation clips when the voltage climbs to about 2.4 volts. If it takes more than about 5 minutes to achieve this voltage, there is something wrong with the battery, so do not use it.

After the depassivation clip is disconnected, the voltage will immediately climb, but will actually take several minutes after the load has been removed before it will climb all the way up to its normal voltage for a new battery, about 3.68 to 3.70 volts for standard temp batteries, about 3.85 to 3.90 volts for the round, low temp batteries. Even if this higher voltage has not yet been achieved, it is still okay to install the batteries into the Tracers; the circuit will normally run on voltage as low as 2.4 to 2.5 volts. Always run the "Test Tracer" routine from the software right after installing a new battery. When it first powers up, the Tracer will be in a high power mode. Once the Tracer has been programmed and read (which is what the "Test Tracer" function does), it will go into a very low power mode and consume very little power from the battery for storage until the Tracer is used next.

At Mesa, we use an internal acceptance specification of 3.50 volts or better, under a 1500 Ohm load to determine whether a battery is "good". The reading should be taken about 15 seconds after applying the load. If the voltage goes down but then up when the load is applied, then the battery was passivated. How much it goes down and then how far it comes up is determined by how passivated the battery was. If the voltage just goes down steadily when the load is applied, then the battery was not passivated.

Back To The Top

Is Mesa Laboratories ISO certified?

Mesa's Medical Division is qualified to the ISO 13485/13488 standard and is registered as a Medical Device manufacturer with FDA, under registration number 1720309 and with Canada (CMDR). Appropriate portions of this rigorous quality control system are also utilized by the DataTrace Division.

The DataTrace Compliance and Certification (form 339) certifies not only the Software Validation but also the FDA-based quality system for DataTrace and the NIST-traceable calibration system. Unlike an ISO certificate, it gives more detail about our quality system. Our Quality Assurance Manual is designed to substantially comply with all the criteria of ISO 9001. This document is revised to keep it current with new products, changing requirements, and technologies.

Back To The Top

Is DataTrace equipment rated Intrinsically Safe and by which agencies?

All DataTrace Tracers are certified intrinsically safe by CSA (Class I, Groups A, B, C, and D; Class II, Groups E, F, and G; Class III; Enclosure 4) and by LCIE in France (EEx ia IIC T4). MPIII Tracers are certified intrinsically safe by LCIE (EEx ia IIC T4). The LCIE certification complies with EN 50014(1992), NF EN 50014(1993), EN 50020(1994), NF EN 50020(1995). DataTrace PC Interfaces are not used in hazardous environments and therefore are not certified intrinsically safe.

Back To The Top

Does DataTrace equipment carry a CE mark?

Yes, all of our DataTrace equipment (Tracers and PC Interfaces) carry the CE mark. The specific requirements relate to the following EC Standards or Norms for Electromagnetic Compatibility: EN 55022 and EN 50082-2.

Back To The Top

Temperature Tracer:

Back To The Top

Can I use boiling water (or an ice bath) to verify my Tracer?

Assuming that proper care is taken in the preparation of these processes, it is possible to verify Tracers in these mediums, but it can be difficult to obtain the stability and uniformity required using "common materials".

First it must be recognized that water boils at 100°C and freezes at 0°C only at sea level. Second, the water used must be ultra pure. Impurities will change both the boiling point and freezing point observed. Finally, the higher in altitude (above sea level) that the verification is performed, the lower the boiling point of water. All of this suggests that a well calibrated reference standard should be used to verify the actual target temperature even in these "known" environments.

The vessel used should be one that is more stable and uniform than a pot on a hotplate. A liquid medium that is not well stirred or isolated from the outside environment will be neither stable nor uniform. Furthermore, the bottom, sides and top can be at significantly different temperatures than the middle.

Back To The Top

Can the Tracer probe tip be made shorter than 0.9" (1.0" for MPIII)?

Physically, it is possible to make the tip 0.5 inches (13mm), less than that creates problems with welding the probe tube, and also, the installation of the sensor into the probe tip, especially the RTD.

Of greater concern is the thermal affect caused by the base on the temperature observed by the sensor. In the old Micropack (both MPII and FRB), tests found in certain environments that the base acted to influence the sensor's temperature readings. Obviously, this depends on the environment and how the Tracer is fixed in the product/process, but there was an observable impact in many situations.

Obviously, the MPIII is much smaller and less of an impact would be expected, but we expect an observable impact in certain situations where the probe tip is less than one inch in length.

Back To The Top

Can the MPIII Tracer be used in a microwave?

Our experience with MPII and FRB Tracers in a microwave environment has been mixed, some good, some bad. In general, the successes have been with applications where the product is homogeneous, with high fluid content, and where the Tracer is completely inside the product. An example of a good application would be pate. An example of a bad application would be pasta drying.

Microwave impacts on MPIII should be similar to previous Tracer versions, with one significant exception - the windows on the side of the body. These become windows that provide direct access for the microwaves to the electronics. This would be bad. If the application were to be tried, the windows must be shielded, either with a cover or at least by positioning them to face away from the microwave source.

Back To The Top

What is the difference between a thermocouple, an RTD, and a Thermistor?

A thermocouple is based on the principle that an electromagnetic force (emf) is generated when heat is applied to the junction of two dissimilar metals (sensing junction). At the other end of the wires, usually as part of the input instrument, is another junction, called the reference junction. The temperature is inferred based on the emf difference between the sensing junction and the reference junction, which is at a known temperature.

Advantages of thermocouples:
Low cost, rugged, small size/fast response, wide temperature range, and reasonably accurate.

Disadvantages of thermocouples:
Weak emf signal, calibration affected by temperature gradients and material contaminants, affected by electrical interference, calibration drifts following calibration.

An RTD is based on the principle that the electrical resistance of a metal increases as its temperature increases. The RTD sensing element consists of pure metal (frequently platinum) and shows a small positive, linear change in resistance per degree of temperature change.

Advantages of RTDs:
High repeatability and stability, high accuracy over wide range, rapid response, and small sensor size.

Disadvantages of RTDs:
High cost, fragile, susceptible to emf interference and self-heating, not suitable for extremely high temperatures.

A thermistor is a thermally active resistor composed of metal oxides normally encapsulated in epoxy or glass. A typical thermistor shows a large negative, nonlinear change in resistance per degree of temperature change. A thermistors resistance drops dramatically and non-linearly with temperature.

Advantages of thermistors:
Inexpensive, long-term stability, high accuracy, temperature coefficient is greater than metals, rugged, small size/fast response, high sensitivity, senses at a single point, and requires simple circuitry.

Disadvantages of thermistors:
Temperature resistance curve is non-linear, relatively narrow range, and not suitable for high temperatures.

Back To The Top

HUMIDITY TRACER:

Back To The Top

How can I clean the humidity sensor on my Humidity Tracer?

The short answer is that the Humidity Tracer's sensor can not usually be cleaned. The sensor's manufacturer does not recommend attempting to clean the sensor, either chemically or by any form of scrubbing as it is likely to cause damage. The humidity sensor has a moisture-sensitive polymer on a ceramic substrate. These components are extremely thin and are sensitive to abrasion. Damage to, or removal of the polymer will destroy the sensor. At best trying to clean the sensor will worsen the affects of the contamination.

The best approach if you believe the Tracer's sensor has become contaminated is to perform a humidity verification. If the calibration is determined to be out of specification, either do a field calibration on the existing sensor replace the Tracer's sensor with a new sensor, or return the Humidity Tracer to the factory for a refurbishment.

Back To The Top

Why are the temperature values of my Humidity Tracer different from my Temperature Tracer?

The environmental temperature recorded by a temperature-only Tracer and the temperature value recorded by a Humidity Tracer can be significantly different. And that is expected. The primary purpose of a temperature-only Tracer is to measure the environment next to the sensor; regardless of what that environment is (e.g., air, liquid, meat, etc.) The temperature sensor's primary purpose in a Humidity Tracer is not to measure the environment; it is to measure the temperature of the humidity sensor. That is the meaning of "temperature compensation". The compensation corrects the humidity reading for the actual temperature the humidity sensor is experiencing - the temperature of the humidity sensor. The environmental temperature is not relevant to the humidity reported.

This is why the temperature reported in a changing environment will be different between a Humidity Tracer and a Temperature-only Tracer. As the temperature in the environment stabilizes, the temperature readings will become closer, eventually becoming the same.

Again, keep in mind that the primary purpose of the Humidity Tracer is to provide accurate humidity values and the primary goal of the Temperature Tracer is to provide accurate temperature values. The methods of accomplishing these goals are not the same for both types of Tracers, as indicated above.

Back To The Top

PRESSURE TRACER

Back To The Top

Can my Pressure Tracer work in a vacuum?

The answer is your Tracers:

  • Can work in a vacuum
  • Yes our Tracer can tell them how much vacuum is pulled as long as they know what the pressure was BEFORE they pulled the vacuum.

All you need to remember is that your Tracer collects ABSOLUTE pressure. In absolute pressure there is no negative pressure, it starts at 0 PSIA and goes up from there. If you want to know how much of a vacuum is pulled you need to know what the pressure was before the vacuum was pulled and what the pressure fell to after you pulled the vacuum. For example, normal atmospheric pressure (at sea level) is 14.7 PSIA. If you pull a vacuum and the Tracer reads 6.0 PSIA, then you have a "vacuum" that is at 6 PSIA.

Normally, when people think of a vacuum they think negative numbers which is typically measured in gage pressure (PSIG). In Gage pressure the ambient pressure (14.7 PSIA at sea level) is 0 PSIG. Using the example above, where you measure 6.0 PSIA, if it were measured in gage it would be -8.7 PSIG (14.7-6.0=8.7).

Back To The Top

How can I clean the pressure sensor on my Pressure Tracer?

Cleaning of the pressure sensor is possible if extreme care is employed. Under no circumstances should swabs, picks, or any other instrument or tool be used in contact with the diaphragm. The diaphragm should never be touched; even the slightest pressure will adversely impact the device and its accuracy.

To start the cleaning process, carefully remove the protective cap by unscrewing it from the probe base.

  • For any coating or process residue adhering to the diaphragm, it is best to soak it off before it dries.
  • For cases where the coating or process residue adhering to the diaphragm has already dried, let the Tracer soak in warm water with a mild detergent until the contamination softens. Then rinse it off with a LOW PRESSURE water stream.
  • In severe cases of contamination, a solvent may be necessary. Use a solvent for the specific contamination material. Since the diaphragm is stainless steel, most solvents will not affect it. If you have a question regarding a specific solvent, contact the factory.

It is critically important that the diaphragm is not touched, scrubbed, picked at, pushed on or probed. This can cause permanent damage.

Back To The Top

How quickly does my Pressure Tracer respond to pressure changes?

The Pressure Tracer' pressure sensor responds to changes in pressure almost instantly (< 1 second). The observed pressure will be accurate providing that the Pressure Tracer is in an isothermal (stable temperature) environment. If the temperature is changing, the observed pressure can be in error by as much as two (2) PSIA.

Back To The Top

Why are the temperature values of my Pressure Tracer different from my Temperature Tracer?

The environmental temperature recorded by a temperature-only Tracer and the temperature value recorded by a Pressure Tracer can be significantly different. And that is expected. The primary purpose of a temperature-only Tracer is to measure the environment next to the sensor; regardless of what that environment is (e.g., air, liquid, meat, etc.) The temperature sensor's primary purpose in a Pressure Tracer is not to measure the environment; it is to measure the temperature of the pressure sensor. That is the meaning of "temperature compensation". The compensation corrects the pressure reading for the actual temperature the pressure sensor is experiencing - the temperature of the pressure sensor. The environmental temperature is not relevant to the pressure reported.

This is why the temperature reported in a changing environment will be different between a Pressure Tracer and a Temperature-only Tracer. As the temperature in the environment stabilizes, the temperature readings will become closer, eventually becoming the same.

Again, keep in mind that the primary purpose of the Pressure Tracer is to provide accurate pressure values and the primary goal of the Temperature Tracer is to provide accurate temperature values. The methods of accomplishing these goals are not the same for both types of Tracers, as indicated above.

Back To The Top

Related Items

DataTrace Products

Contact Us