Mesa Home>Nusonics>Concentration Analyzer>Model 87 Lab Analyzer>Specifications

Model 87 Lab Analyzer Specifications

Performance

Range Sound Velocity: 500 - 2500 meters/sec (any acoustically transparent liquid)
Temperature: 0 to 100 ° C (5 - 365 ° F) Standard
Resolution Sound Velocity: ± 0.01 meters/sec
Temperature: ± 0.01 ° C
Repeatability Sound Velocity: ± 0.10 meter/sec
Temperature: ± 0.05 ° C
Response Time 2 Seconds
Accuracy Sound Velocity: ± 0.20 meter/sec
Temperature: ± 0.10 ° C
Concentration: Application Dependent, typical values are ... Acids and Caustics: ± 0.05 Wt. %
Salts: ± 0.1 Wt. %
Organic liquids: ± 0.1 to 0.2 Wt. %
Oils in water: ± 0.2 Vol. %
Sugars: ± 0.1 °Brix
Calibration Stability 3+ Years, based on 0.003 mils/year corrosion rate of wetted materials.
Certifications N.I.S.T. traceable certificates available for certain applications.

Power Requirements

Input Voltage 115 (+/- 10%) volts AC
Optional Input Voltage 230 (+/- 10%) volts AC
Frequency 50-60 hertz

Transmitter Power Consumption

Standard 35 watts
With Heater 200 watts @ 115 VAC
  250 watts @ 230 VAC

Output Signals

Interface Dual 4-20 mA into 600 Ohms max each
Fault Indication Form C Relay
High/Low Set Points Form C Relay
Data Link RS-232, RS-485 optional

Enclosure Dimensions

NEMA 4X 15.3" (h) x 13.3" (w) x 8.3" (d)
NEMA 7 21.9" (h) x 15.9" (w) x 10.8" (d)

Weight

NEMA 4X 20 lbs. (9.1 kg)
NEMA 7 114 lbs (51.7 kg)

Operating Temperature Range

Standard Transmitter 0° C to +50° C (32° to 122° F)
With Heater -60° C to 50° C (-76° F to 122° F)
Standard Transducer -40 C to 150 C

Product Material

Display Liquid Crystal (LCD)

Related Items

Theory of Operation

Tested Substances

Sulfuric Acid Analysis

Sodium Hydroxide Analysis

Sodium Chloride Analysis

Fruit Juice Applications

Rolling Oil Solutions

Ammonium Hydroxide Analysis

Other Applications

(click to inlarge)
Model 87