DataTrace Thermal Barriers

If the process you want to monitor reaches temperatures above 140ºC (the upper operating limit of DataTrace Micropack III data loggers) you need to thermally protect the electronics and battery from damage. For these applications you can choose from 10 different DataTrace Thermal Packs, depending on your particular application. These thermal barriers are designed to insulate the logger body from the high temperatures, while allowing the temperature probe to extend out into the environment. All of the barriers work well in a dry environment and most will work in liquid environments, such as hot oil baths. In addition to high-temperature applications, the thermal barriers will protect the data logger battery in ultra-low temperature environments.

Several DataTrace Thermal Packs are available, depending on your particular needs.


Business Applications

  • Extend the use of your DataTrace data loggers to high temperature processes, getting more from your capital investment.
  • Eliminate the need for unstable and cumbersome thermocouple-based measurement systems, saving you time and money.
  • Monitor additional manufacturing processes, optimizing your productivity.

 

Food Applications

  • Baking Oven Monitoring
  • Frying Oil Temperature

 

Pharmaceutical Applications

  • Depyrogenation Ovens
  • Ultra-low Temperature Freezers

 

Industrial Applications

  • Textile and Carpet Processing
  • Powder Painting Monitoring

Model 2150 - Thermal Barrier

For MPRF loggers, Model 2150 holds one MPRF logger.

 

 

Model 2150 - Specifications

Size

Cylinder, 50 mm D 152 mm L

Loggers

One MPRF Data Logger

Insulation

PTFE

Use

Dry Only

Typical Dry Performance

Temperature Maximum Time

250ºC/482ºF

27 minutes

350ºC/662ºF

20 minutes

400ºC/752ºF

17 minutes

Model 3000 - Moderate Temperature

Designed for either dry or liquid environments, the Model 3000 is an economical choice to protect a single logger in moderate temperature environments.

 

Model 3000 - Moderate Temperature Thermal Pack

Model 3000 - Specifications

Size

Cylinder, 75 mm D 120 mm L

Loggers

One MPIII

Insulation

PTFE

Use

Dry or Liquid

Typical Dry Performance

Temperature Maximum Time

250ºC/482ºF

59 minutes

350ºC/662ºF

42 minutes

400ºC/752ºF

38 minutes

Model 3080 - Multi Logger

With the lowest cost per measurement channel, the Model 3080 is designed to protect three Micropack III data loggers in moderate temperature environments.

 

Model 3080 - Multi Logger Thermal Barrier

Model 3080 - Specifications

Size

Rectangular Box

95 mm X 82 mm X 46 mm

Loggers

Three MPIIIs

Insulation

PTFE, Carbon

Use

Dry Only

Typical Dry Performance 

Temperature Maximum Time

250ºC/482ºF

36 minutes

350ºC/662ºF

27 minutes

400ºC/752ºF

24 minutes


Models 3070/3071 - Single Logger

Single-logger versions of the 3080 with the same temperature performance characteristics.  Model 3070 holds a single-probe logger. Model 3071 holds a dual-probe logger.

 

3071 Rectangular Thermal Barrier Dual Probe

Models 3070/3071 - Specifications

Size

Rectangular Box

64 mm X 86 mm X 44 mm

Loggers

One MPIII

Insulation

PTFE, Carbon

Use

Dry Only

Typical Dry Performance 

Temperature Maximum Time

250ºC/482ºF

36 minutes

350ºC/662ºF

27 minutes

400ºC/752ºF

24 minutes

 

Models 3140, 3141, 3200 & 3400 – High Temperature

For the ultimate in protection, choose from three different 2-inch diameter vacuum dewar thermal barriers. The largest, Model 3400, will protect your loggers for several hours at temperatures above 350ºC.

 

High Temperature Thermal Barriers

Model 3140, 3141, 3200, 3400 - Specifications

Model # Model 3140 Model 3141
Model 3200 Model 3400
Size

Cylinder, 50 mm D 140 mm L

Cylinder, 35 mm D 140 mm L

Cylinder, 50 mm D 200 mm L

Cylinder, 50 mm D 400 mm L

Loggers

One MPIII

One MPIII

One MPIII

One MPIII

Insulation

Vacuum Dewar, PTFE

Vacuum Dewar, PTFE

Vacuum Dewar, PTFE

Vacuum Dewar, PTFE

Use

Dry or Liquid

Dry or Liquid

Dry or Liquid

Dry or Liquid

Temperature Maximum Time Maximum Time Maximum Time Maximum Time

250ºC/482ºF

109 minutes

87 minutes 195 minutes 380 minutes

350ºC/662ºF

77 minutes 54 minutes 135 minutes 240 minutes

400ºC/752ºF

68 minutes 42 minutes 120 minutes 205 minutes

Model 3175 – Slim Line

If space is at a premium, such as in depyrogenation ovens, the 1.2 inch diameter Model 3175 vacuum dewar is an excellent choice for the ultimate protection of your Micropack III data loggers.

Model 3175 – Slim Line Thermal Barrier

Model 3175 - Specifications

Size

Cylinder, 30 mm D 175 mm L

Loggers

One MPIII

Insulation

Vacuum Dewar, PTFE

Use

Dry or Liquid

Typical Dry Performance

Temperature Maximum Time

250ºC/482ºF

93 minutes

350ºC/662ºF

61 minutes

400ºC/752ºF

53 minutes

 